AM5000 Quantitative Fluorescence Polymerase Chain Reaction (PCR) System
Cat. No. ACB40001
Product Name | AM5000 Quantitative Fluorescence Polymerase Chain Reaction (PCR) System |
Cat. No. | ACB40001 |
Sample Capacity | 96 orifice、12*8 coupling、96*0.2ml singlet (transparent bottom) |
Dynamic Range | 1~1010 copies |
Excitation Wavelength | 300-800 mm |
Detection Wavelength | 500-800 mm |
Five-channel Applicable Dye | F1:FAMSYBR Green I; |
F2:VICHEXTETJOE | |
F3:Cy3NEDTAMRA | |
F4:ROXTEXAS-RED; | |
F5:Cy5; | |
Module Operating Temperature Range | 4ºC~105 ºC (Minimum setting scale: 0.1 ºC) with SOAK Cryopreservation function |
Heating and Cooling Rates | 4.5 ºC /s (max) |
Temperature Control Accuracy | ≤ ±0.1 ºC |
Temperature fluctuation | ≤ ±0.1 ºC |
Temperature uniformity | ≤ ±0.3 ºC |
Temperature Control Mode | BLOCK, Simulation Tube Mode (According to automatic dosing amount) |
Dosing Volume | 10-100 μL |
Gradient Temperature Range | 1 ºC ~36 ºC |
Thermal Lid Temperature Range | 30 ºC ~110 ºC( Adjustable, Default 105 ºC, Automatic thermal lid) |
Fluorescence Intensity Detection Repeatability | 5% |
Scan Mode | Full plate scanning, Specify line scan |
Programming | Each program can set up 20 temperature control segment; the maximum number of cycles is 99. |
Operation Mode | Continuous operation |
Scan Time | Approximately 5.5s (F1 full plate 96 orifice scan) |
Features | Absolute quantification, relative quantification, SNP analysis, automatic data analysis, HRM. |
Gradient and automatic gain, user-defined dye (combined analysis of multiple results and files.) | |
Operating System | Application Software:Microsoft:Windows XP/vista/windows 7/windows 8(Compatible with 64-bit) |
Microsoft:Excel2000/2002/2003/2007 | |
Minimum Computer Configuration | RAM 1G,Hard drive 80G,CPU 1.5GHz |
Input Power | 100-240V 50/60Hz 600W |
Dimensions (L*W*H) | 430mm*395mm*352mm |
Weight | 28 Kg |
Interface | Comes with USB converter and optional purchase of RS232C converter, Bluetooth wireless converter (instruments with 4 or more channels will be equipped with three types of converters). |